X-Ray eye info from old web site - not sure of validity:
******* X-RAY EYE *******
Software is in zircon: /home/user1/xrayeye/dt6
Yorick routines are in: xrayeye/routines
Data (images) are stored in: xrayeye/data
To run the program login as user1 go to xrayeye directory and run: gl dt6
commands:
TV,window,cmax=,cmin=,channel=0
Frame grabber can handle four tv inputs (4 channels).
TV,0 // to make tv plot in window 0.
++++++++++++++++++++++++++++++++++++++++++++++++++++++++
WARNING: DON'T USE CTRL-C TO STOP ANIMATION - USE RETURN
++++++++++++++++++++++++++++++++++++++++++++++++++++++++
NTV,0,50 // to make average of 50 frames.
This doesn't do much unless there is random noise.
These two functions are included in "newtv.i"
savimg,name,comment
Save img in a file with name and comment string
This function is included in "imgsav.i"
Other commands:
ccd_bw,lowerlevel,upperlevel
set the voltage levels on the framegrabber. Use help,ccd_bw
for more info.
ccd_read,image,channel,framing
this open camera if needed and reads one image.
TROUBLESHOOTING:
If the computer was rebooted mount /flinthome/
TECHNICAL NOTES:
- FIELD OF VIEW: 2.8 X 2.1 mm
- RESOLUTION: 4.2 microns per pixel
- CAMERA SHUTTER SPEEDS: 0 1/60
1 1/125
2 1/250
3 1/500
4 1/1000
5 1/2000
6 1/4000
7 1/10000
- GAIN SETTING: |