Diamond Monochormator ion pump operation instructions

On the controller, ion pump #1 is the larger pump attached to the vacuum chamber and ion pump #2 is the smaller pump hooked-up to the intermediate stage of the rotary seal.

Starting
  • Use the Pump Select button to select which pump to operate
  • Press Amp/Volt to display voltage
  • Press HV ON/OFF for the pump you want to start
    (Displayed voltage should be increasing rapidly.)
    After about 30 sec., press Start Prot to engage the Protection Mode.
    Current should fall in the 10-4 A

Stopping
  • Use the Pump Select button to select which pump to operate
  • Press Start Prot to disengage the Protection Mode
  • Press HV ON/OFF for the pump you want to stop


Protection Mode trip
If the pump stops due to a Protection Mode trips shortly after turning it on, the HV light (amber) will flash. This most likely means that the pump cannot start because pressure is still too high.
In order to re-start the pump, you first have to:
  • Press Start Prot to disengage the Protection Mode (green light should be off)
  • Press HV ON/OFF so that the HV light is off