[6/25 7:38 PM] Narayanan, Suresh
lambda250k with all the modes with and without trigger has been tested quite thoroughly from spec.
All the shutter functionality for detector control is like 8idi now, can run with or without shutter
for fast and slow measurements, maybe helpful for some rad sensitive samples, documented it
reasonably well in paper, not as good as your docs
Att. 1 is from Suresh detailing the Softglue work done in May.
[7/113/21 10:04 PM] Narayanan, Suresh
Sure. Cabling was a lousy job that day with the DP person, trying to do it when i was out the door.
IOC is running on the DP PC on the rack.
[7/13/21 10:05 PM] Narayanan, Suresh
One thing I did not check was the physical shutter opening. The status of the fast shutter.ui
screen works like before so I assume it should be good.
New softglue: https://logbook.xray.aps.anl.gov/8id/8-ID/787
PO/Upgrade notes from Suresh, Xspectrum and DP staff: https://logbook.xray.aps.anl.gov/8id/8-ID/773
See new spec macros:
xpcs_ad_nofpga_lambda
use_shutter_mode_regular
use_shutter_mode_burst
From batches/2021/May2021 lambda.do
##Usage: padtimescan ndata xtime [period]
# For shuttered mode, choose LAMBDA_OPERATING_MODE = 1 and EXT_TRIGGER=2
# For ext. trigger mode 0, 1, choose LAMBDA_OPERATING_MODE = 0
# zero deadtime (ZDT)
LAMBDA_OPERATING_MODE = 3; ##0-1 bit ZDT, 1-6-bit ZDT, 2 - 12 bit ZDT, 3 -24-bit
mode
EXT_TRIGGER=2
att 0
ARun_number++
batch_name=sprintf("A%03d_delete",ARun_number)
#padtimescan 50000 1/24000.0
padtimescan 50 .1 .55
p INFOFILE
EPICS screen now on 8ID-E Engineering screen
Click on AreaDetector in att. 2.
The full screen appears as shown in att. 3.
Pushing the "less button" at the bottom of the screen gives a smaller, narrower window.
How to start IOC
xspadmin@xspserver:~/bin$
xspadmin@xspserver:~/bin$ ./DP_LAMBDA250K_1.sh status
DP_LAMBDA250K_1 is running (pid=13422) in a screen session (pid=13421)
xspadmin@xspserver:~/bin$ ./DP_LAMBDA250K_1.sh stop
Stopping DP_LAMBDA250K_1 (pid=13422)
xspadmin@xspserver:~/bin$ ./DP_LAMBDA250K_1.sh
Usage: DP_LAMBDA250K_1.sh {start|stop|restart|status|console|run|medm|caqtdm}
Additional options:
DP_LAMBDA250K_1.sh start {screen|procServ|ps|shell}
xspadmin@xspserver:~/bin$ ./DP_LAMBDA250K_1.sh status
DP_LAMBDA250K_1 is not running |