8-ID
  8-ID Beamline logbook mirror  Not logged in ELOG logo
Message ID: 809     Entry time: Sat Sep 25 11:45:41 2021
Author: Eric Dufresne 
Type: 8-ID-I 
Category: Vacuum 
Subject: 8-ID-I televac gauges 
Re: Logging the incident on unknown att1 drop
Dufresne, Eric <[email protected]>
Sat 9/25/2021 12:07 PM
To: Zhang, Qingteng <[email protected]>; Narayanan, Suresh <[email protected]>
2 attachments (604 KB)
IMG_1293.jpg; Televac.png;

Qingteng, it was a lucky check with the filters, but maybe I suspected this possibility.

I got fooled also by a picture from you about the downstream
Televac (see pic). Note that there are 3 Televacs: TV3, TV4, and TV6.
The vacuum in TV6 was set on input 2 instead of 4 which can be toggled by the
Up/down arrows in blue.

I include here a picture of the second unit in EPICS also.
I think SP4 is the downstream high vacuum. I am not sure
what's the upstream flight path in EPICS, but it is on
#1 on TV6 in your picture.

Eric
Attachment 1: Televac.png  42 kB  Uploaded Sat Sep 25 12:46:26 2021  | Hide | Hide all
Televac.png
Attachment 2: IMG_1293.jpg  529 kB  Uploaded Sat Sep 25 12:53:34 2021  | Hide | Hide all
IMG_1293.jpg