8-ID
  8-ID Beamline logbook mirror  Not logged in ELOG logo
Message ID: 827     Entry time: Tue Feb 1 10:38:23 2022
Author: Eric Dufresne 
Type: 8-ID-I 
Category: Important Log Books 
Subject: Logbook notes for 2022-1 for 8-ID-I operation 
https://logbook.xray.aps.anl.gov/8id/8-ID/596 Important distances 2018-3
https://logbook.xray.aps.anl.gov/8id/8-ID/674 Lens positions for 1 and 2D focusing
https://logbook.xray.aps.anl.gov/8id/8-ID/701 2020-1 notes
https://logbook.xray.aps.anl.gov/8id/8-ID/721 2020-2 notes
https://logbook.xray.aps.anl.gov/8id/8-ID/744 2020-3 notes
https://logbook.xray.aps.anl.gov/8id/8-ID/761 2021-1 notes
https://logbook.xray.aps.anl.gov/8id/8-ID/761 2021-2 notes
https://logbook.xray.aps.anl.gov/8id/8-ID/827 2022-1 notes

https://anl.app.box.com/folder/130409499261?s=3t71weuketeea2s3xnvoo59tcclxk3dw Handwritten notes

8-ID-I ESAF 2022-1 https://beam.aps.anl.gov/pls/apsweb/esaf0001.start_page?i_attrib=250714

8-ID-E ESAF 2022-1 https://beam.aps.anl.gov/pls/apsweb/esaf0001.start_page?i_attrib=251426

1/27/2022 Report of issue with power cycle on 8ID-A IOC (MaxV card doesn't restore PVs)
See fix https://logbook.xray.aps.anl.gov/8id/8-ID/826

2/1/2022 First day of beamtime 8-ID-I

2/15/2022 Gravity fall water trip/problem
https://logbook.xray.aps.anl.gov/8id/8-ID/830

2/28/2022 8ID-I hutch door repair impacting all of 8ID
https://logbook.xray.aps.anl.gov/8id/8-ID/831

3/22/2022 First day of beamtime in 8ID-E WA-XPCS
I fixed the autoShutter database. For some reasons, it had lost its permit.

Tough day in 8-ID-E. It started with the camera switch being broken.
Then I had no x-rays anywhere, pind3, pind4, etc... so I enventually identified the shutter.
What I should have done was to test the shutter; you can view the shutter open
and close on the viewport.

I finally identified it as a EPICS issue at 5h32. The SG MZ3 unit had losts its
entries. This is described here: https://logbook.xray.aps.anl.gov/8id/8-ID/834
Suresh and I are still thinking there is another problem with the fast shutter EPICS screen
which remains inverted.

4/19/2022

https://logbook.xray.aps.anl.gov/8id/8-ID/838 Displaced the mirror down by 0.5 mm to get back
my flux at 7.35 keV.