Message ID: 34     Entry time: Fri Feb 17 17:03:36 2006
Author: Alec 
Type: 8-ID General 
Category: Detectors 
Subject: Scintillators 
Data on the bicron scintilator in use at IMM-CAT

Settings on electronics for 7.66 keV:
High Voltage Supply: Ortec 456: 700V
Amplifier: Ortec 590: Settings: fine gain: 0.5 +7 tics
course gain: 20
SCA: Canberra SCA 2030: Settings: Window = 5.20
LL = 2.60

Input pos.
The effective dead time of this setup is around 6 micro-sec, a
more accurate number should be available soon ...