  813   Tue Oct 12 15:10:24 2021 Eric Dufresne8-ID-IDetectorsDetails of how 8-ID-I is connected with Joergers and softglue SG-B and IC2 set up

We have two Nova V2F in 8-ID-I.
The left Nova V2F in att. 1 has PDBS , the right one has pind1, pind2, pind3 (IC1), pind4.
The two outputs on the V2F goes:
1) to the Joerger EPICS scaler
2) to the Joerger Display above bronze, model QVSC.

In the middle of the 8-ID-I rack, on the right, there is a board with 2 rows of 8 signals, these are
TTL-NIM converters. The inputs takes the output of the V2Fs and send them on the output to the
Joerger QVSC. So on this board, Input 4 is pind4, and input 5 is pdbs.

Finally Suresh has configured SoftGlue unit B here referred as SG-B input channel 6
to record the transmission detector. In vacuum it would be pind4, but Qingteng
wants to record the second ion chamber instead, so today we wired this.
The SG-B panel is shown in att. 4.


The SR570 PDBS unit (#5) is not responding. Using instead spare unit from Sec 26. Everything works,
except that the spare unit is not connected serially so at the moment only pind3 can be controlled in EPICS.
We set the same gain at 10 nA/V. The signal on IC2 is about 80% of IC1. I think I don't need to check the
impact of the T (kind of difficult to remove too).

ED
Attachment 1: Nova-V2F.png
Nova-V2F.png
Attachment 2: JoergerQVSC.png
JoergerQVSC.png
Attachment 3: 8idiSGB-Inputs.png
8idiSGB-Inputs.png
Attachment 4: SGBpanel.png
SGBpanel.png
  812   Fri Oct 8 12:11:48 2021 Eric Dufresne8-ID-IDetectorsCan one use the MCS to see 147 Hz
I had a question to Suresh.

There is the direct method, that is plug the signal from the
pin diode after the SR-570 and look at the voltage on a scope.
On pind1, the signal was roughly 1.5V on 10/5/21 for 1D focusing at 11 keV,
so high ~like 7x10^11 ph/s, at 200 uA/V. We observed 0.1V peak to peak.
https://anl.box.com/s/qj29qzajsxgo0gzgjddy1wsqaw73yzg1

We have handy a Tektronic 3032 DPO scope with 300 MHz analog bandwidth and 2.5 GS/s with a NIC.
Can it be interfaced in EPICS?
Looks like it might https://epics.anl.gov/modules/instrument/tds3000/index.html.

Yes it can.
Re: Your TDS 3000 EPICS device support
DiMonte, Nicholas <[email protected]>
Mon 10/11/2021 12:02 PM
To: Dufresne, Eric <[email protected]>
Hi Eric,

There is no script, the waveforms can be read via the pvs which are 10k words.
SDDS basically reads these waveform pvs.

Thanks,
Nicholas





I'm also thinking of using the MCS on the 8-ID-I crate to determine the flux variations
and take an FFT from the signal.

How is the Joerger MCS set up, does it takes TTL or NIM pulses?
I forget how the Nova V2F are set up, do you remember if they output TTL or NIM pulses?
[1:34 PM] Narayanan, Suresh
NOVA does TTL only. Joerger MCS or Struck MCS, it takes TTL as well.
  811   Mon Oct 4 10:01:07 2021 Eric Dufresne8-ID-IImportant Log BooksLogbook notes for 2021-3 for 8-ID-I operation
https://logbook.xray.aps.anl.gov/8id/8-ID/596 Important distances 2018-3
https://logbook.xray.aps.anl.gov/8id/8-ID/674 Lens positions for 1 and 2D focusing
https://logbook.xray.aps.anl.gov/8id/8-ID/701 2020-1 notes
https://logbook.xray.aps.anl.gov/8id/8-ID/721 2020-2 notes
https://logbook.xray.aps.anl.gov/8id/8-ID/744 2020-3 notes
https://logbook.xray.aps.anl.gov/8id/8-ID/761 2021-1 notes
https://logbook.xray.aps.anl.gov/8id/8-ID/761 2021-2 notes

https://anl.app.box.com/folder/130409499261?s=3t71weuketeea2s3xnvoo59tcclxk3dw Handwritten notes

8-ID-I ESAF 2021-3 https://beam.aps.anl.gov/pls/apsweb/esaf0001.start_page?i_attrib=247172

10/4/2021 Alignment 10.945 keV

8ID-E mono was at 10.945 keV.
https://anl.app.box.com/file/868578355997 Notes
[9:22 AM] Dufresne, Eric
Flux is the same as on 9/21, so mirror and diamx are in good shape. What we found last fiscal year is that when
the operation mode doesn't change, the source point returns roughly to the same position from week to week.
Flux is 1.3x10^12 ph/s with 250x250 si1 slit, mono at 10.945 keV as last alignment on 9/21/21.
<https://teams.microsoft.com/l/message/19:[email protected]/1633357336064?tenantId=0cfca185-25f7-49e3-8ae7-704d5326e285&amp;groupId=4c5a5c95-cec4-4898-b2db-37615c5a8ceb&amp;parentMessageId=1633355943169&amp;teamName=DYS_General&amp;channelName=8-ID-I Ops&amp;createdTime=1633357336064>

10/12/2021 Alignment unchanged in 8-ID-I
https://anl.box.com/s/68gb05v652pquhhom7sagzju4xz1jm70 Notes.
[8:54 AM--Teams 8ID-I OPs] Dufresne, Eric
Flux in 8-ID-I pind1 is same as last week, so I'm not touching the mirror.
Flux on pind1 is 7.48x10^11 ph/s in 250 um x 150 um (HxV).
Flux on pind2 is also good. Flux is 3.61x10^10 ph/s, very very close to
last week. I am getting ready now.

https://logbook.xray.aps.anl.gov/8id/8-ID/813 Set up with QZ and SN IC2.
Above describes electronics.
https://anl.box.com/s/gh8zzej4j3a194ys8q43yiotsz9dwu2v Lab notes 10/12/21

10/13/2021

Water leak inside ring
See pictures https://anl.box.com/s/3wec9gzrkx4t0gv9nrv3qdil6tig020y
with PDF file of repair status.
Beam was off after 2h am on 10/13.

10/19/2021
https://anl.box.com/s/mr5f9fza0fpeg2bofrrtnovk6aus8gyq Notes

Flux on pind1 was 2% below that of 10/13/21.
Flux on pind2 was right on.
Flux on pind4 was 2.2x10^10 ph/s

11/02/2021
ALignment at 7.35 keV
8ID-I Wei Chen focus at sample
8ID-E Shpyrko
https://anl.app.box.com/file/880492191996 My 8ID-I notes

11/9/2021
https://anl.app.box.com/file/882923569293 My 8ID-I notes
Alignment at 10.945 keV.
N=15 Be lenses R=0.2mm 1D

[9:11 AM] Dufresne, Eric
Zhang, Qingteng Strzalka, Joe Narayanan, Suresh Started aligning 8ID
[9:38 AM] Dufresne, Eric
Flux with 250umx250um si1 is 1.08x10^12 ph/s in 8-ID-I. We can align 8ID-E Joe.
&#8203;[9:41 AM] Dufresne, Eric
This is lower than on 10/4/21, 1.31/1.08= 1.21 so 21%. I am not hunting this down Narayanan, Suresh.
&#8203;[10:31 AM] Dufresne, Eric
A lot less signal on pind2 FYI, only 1.44x10^10 ph/s (flux pind2 32493-1355), i..e 40% of 10/12/21.
Didn't see anything obvious. We are only 87% of pind1 from 10/12.
TeamsChatHyperlink

11/15/2021

We determined that the lost flux last week was because we didn't pump the 8-ID-I beamline.

11/17/2021
[9:41 AM] Dufresne, Eric
Strzalka, Joe I think we are happy in 8-ID-I. Fell free to start Jiang, Zhang and Strzalka, Joe.

To summarize, we are in a new mode of operation in Hybrid mode from 324 bunch-mode last week. Flux in 8-ID-I is back
to normal at 1.33x10^12 ph/s in a 250x250 um si1, and 7.93x10^11 ph/s in a 250 (H) x 150 (V) beam on pind1.

<https://teams.microsoft.com/l/message/19:[email protected]/1637163677964?tenantId=0cfca185-25f7-49e3-8ae7-704d5326e285&amp;groupId=4c5a5c95-cec4-4898-b2db-37615c5a8ceb&amp;parentMessageId=1637160445263&amp;teamName=DYS_General&amp;channelName=8-ID-I Ops&amp;createdTime=1637163677964>

[9:45 AM] Dufresne, Eric
Zhang, Qingteng The beam is also in good shape on pind2 with a flux of 3.28x10^10 ph/s in a 15 um H si1hgap.
This is only 10% lower that on 10/19/2021 where it was 3.6x10^10 ph/s. Recall last week it was only 1.44x10^10 ph/s.
I am going to let you take over. I have something at 10am for APS-U, and I don't know what you have in the sample stage.

(1 liked)<https://teams.microsoft.com/l/message/19:[email protected]/1637163932527?tenantId=0cfca185-25f7-49e3-8ae7-704d5326e285&amp;groupId=4c5a5c95-cec4-4898-b2db-37615c5a8ceb&amp;parentMessageId=1637160445263&amp;teamName=DYS_General&amp;channelName=8-ID-I Ops&amp;createdTime=1637163932527>

I did also perform a crlx scan with pind2 with si2 at 150 um (V) by 15 um (H).
This is shown here: https://8id.xray.aps.anl.gov/specplots/8idi/2021/11/tingxu20211117/s00002.svg
The transmission is quite flat, would expect the beam footprint of 0.25 mm so looks good
when sampled with 15 um beam on si2hgap.

12/07/2021
Preparing for 7.35 keV in 8ID-I.
Will focus on sample, equivalent to N=7, R=0.2 mm.
Labelled earlier a lens set with red dot.
I think I located it so that I would use it this time.
Also labelled a second set with a blue dot, set was labelled 4A as well.

12/08/21
Good alignment of 8-ID for 7.35 keV ops. Flux down 13% from Nov 2 alignment.
https://anl.app.box.com/file/893637342395
  810   Fri Oct 1 14:46:20 2021 Eric Dufresne8-ID-IOptics2D focusing in 8-ID-I imaging the pinhole at 10.9 keV with New Be lens order
I made all the spacers in 2020. I tested it at 7.35 keV in 8-ID-E only,
and I claimed that it didn't work in general, this is discussed in
8ID elog: https://logbook.xray.aps.anl.gov/8id/8-ID/692 below.

I didn't write any evidence backing this claim.
I wrote this in my JSR paper for 7.35 keV in the
appendix C. I measured something before my beamtime in 8-ID-E with the WB slits opened
and wrote it in Figure 15 https://journals.iucr.org/s/issues/2020/06/00/gb5109/gb5109fig15.html.

"One would expect to either demagnify the horizontal source (20 µm FWHM) or the pinhole (16.6 µm FWHM).
We observe 28.6 µm FWHM horizontally. We suspect this broadening of the focal spot is caused by the
heat bump of the Si monochromator which is not observed in the vertical."

https://journals.iucr.org/s/issues/2020/06/00/gb5109/gb5109.pdf



Quote:

My calculations and drawings for the new x-ray optics set up
to focus in 2D at 10.9 keV.

Attachment 1 show the new 23 mm spacers required to keep the existing spring at the
same tension as before for other conditions at 7.35 keV. This drawing was put in the shop 1/14/2020.
To focus at 11 keV, we need N=8 R=0.1 mm 2D Be lens.
This is a mount I first introduced here: https://8id.xray.aps.anl.gov/elog/8-ID/616.

I've ordered 4 additional R=0.1 mm lenses last November.
https://apps.inside.anl.gov/paris/req.jsp?reqNbr=G9-319040.1
They were received January 30, 2020. The average dead thickness is 24.5 microns, 4 were deliverd.

My report is also attached, V 1.2, updated from an email I sent in early November.

N.B. Derived in my notes p 32-33 for transmission of lens in book started Sep. 21, 2019
See /home/beams/DUFRESNE/8ID/StrategicPlan/2019/2D-focus-8-ID-I-UsingPinhole for
all the code and Latex documents.

Shop order
Date: Wed, 12 Feb 2020 14:50:43 +0000
From: Argonne Service Desk <[email protected]>
To: [email protected]
Subject: FMS0050724 - TM-Eric Dufresne - 8-ID-I Mount Accessories,
12mm OD x 23mm long with 4mm clearance hole thru length,
132 > PRJ1000854 APS03-132 > PH01 General > A
Parts/Attachments:
1.1 OK 15 lines Text
1.2 Shown 305 lines Tex


On July 2020, Eric found out this scheme doesn't work.
One would have to correct the horizontal focusing.
  809   Sat Sep 25 11:45:41 2021 Eric Dufresne8-ID-IVacuum8-ID-I televac gauges
Re: Logging the incident on unknown att1 drop
Dufresne, Eric <[email protected]>
Sat 9/25/2021 12:07 PM
To: Zhang, Qingteng <[email protected]>; Narayanan, Suresh <[email protected]>
2 attachments (604 KB)
IMG_1293.jpg; Televac.png;

Qingteng, it was a lucky check with the filters, but maybe I suspected this possibility.

I got fooled also by a picture from you about the downstream
Televac (see pic). Note that there are 3 Televacs: TV3, TV4, and TV6.
The vacuum in TV6 was set on input 2 instead of 4 which can be toggled by the
Up/down arrows in blue.

I include here a picture of the second unit in EPICS also.
I think SP4 is the downstream high vacuum. I am not sure
what's the upstream flight path in EPICS, but it is on
#1 on TV6 in your picture.

Eric
Attachment 1: Televac.png
Televac.png
Attachment 2: IMG_1293.jpg
IMG_1293.jpg
  808   Wed Sep 22 10:52:14 2021 Eric Dufresne8-ID-IX-ray Source12 kHz line as well as 20 or 40 Hz.
Re: 12 kHz line
Emery, Louis <[email protected]>
Wed 9/22/2021 12:36 AM
To: Dufresne, Eric <[email protected]>
Cc: Narayanan, Suresh <[email protected]>; Zhang, Qingteng <[email protected]>

We don't see the same lines.
We see in the tune signals from a stripline (searching for high frequencies) a line at 21.5 kHz in y plane,
and lines at 7.0, 18.7, 37 kHz in x plane. But not at 12 kHz.

From rf bpms (searching for lower frequencies) we have 60, 120 and 147 Hz lines.
but none at 20 or 40 Hz.

Louis Emery
Senior Physicist
Advanced Photon Source
Argonne National Laboratory
9700 South Cass Avenue
Argonne, IL 60439

From: Dufresne, Eric <[email protected]>
Sent: Tuesday, September 21, 2021 11:37 PM
To: Emery, Louis
Cc: Narayanan, Suresh; Zhang, Qingteng
Subject: 12 kHz line

Hi Louis,

hope you are doing well.
My colleague Qingteng shared with Suresh and I this data.
He was puzzled with the 20-60 Hz lines which we do not typically see.
I suspect they might be caused by us.

He also noticed something at 12 kHz. Anything from your end at this high a frequency?

Cheers,
Eric
Attachment 1: FFT-8ID-092121-RigakuData.png
FFT-8ID-092121-RigakuData.png
  807   Tue Sep 21 14:27:41 2021 Eric Dufresne8-ID-ISample EnvironmentsNew double Peltier stage for fly scanning stage--updates in sample mounts by QZ
Re: Adaptor Piece
Zhang, Qingteng <[email protected]>
Tue 9/21/2021 12:57 PM
To: Laurence Lurio <[email protected]>; Narayanan, Suresh <[email protected]>
Cc: Dufresne, Eric <[email protected]>; Basnet, Gobind <[email protected]>; George Thurston <[email protected]>
1 attachments (814 KB)
IMG_1262.jpeg;
Dear All:

The new adapter plate is ready. We can now mount up to 12 capillaries at the same time. I put all the pieces on the stainless steel bench in the chem lab next to George's samples.

@Laurence Lurio @George Thurston Should the samples be kept in the 4C fridge?

Thanks,


QZ

Re: Measurement protocol
Zhang, Qingteng <[email protected]>
Sun 9/19/2021 5:23 PM
To: Laurence Lurio <[email protected]>; Narayanan, Suresh <[email protected]>
Cc: Dufresne, Eric <[email protected]>; Basnet, Gobind <[email protected]>; George Thurston <[email protected]>
2 attachments (1 MB)
IMG_1255.jpeg; IMG_1256.jpeg;
Larry:

I did some quick measurements on the blocks and here's what I find:

-- I should be able to fit 3 blocks on the fly stage with <10 mm of extension (see photo attached).
I'll send Ray the drawing for adaptor plate so that he could drop it at the machine shop first thing in the morning;

-- The opening of the block is 10 mm long, meaning anything in the capillary taller than 10 mm will be clipped
by the block (there's almost no clipping at the bottom though), so I would not recommend 25 mm tall of samples (see photo attached);
..............




Quote:
Dana, Qingteng and Don Jensen designed a new double-Peltier stage for
fly scanning with multiple samples to test. We tested one with 5 as shown in att. 7 and 8.

The attachment below are the last settings of the new cell.
Feedback on IN3 of Lakeshore 336, type K. Othe rtype K on IN4.
This uses the new fly scan capability coupled with a new heater block.
Some of the technical motor location are shown in att. 5 (mostly for my own good).

Also ion chamber can be plotted from /home/beams/8IDIUSER/User_XPCS_DATA_Logs/
Attachment 1: IMG_1262.jpeg
IMG_1262.jpeg
Attachment 2: IMG_1256.jpeg
IMG_1256.jpeg
  806   Tue Sep 14 23:44:08 2021 Eric Dufresne8-ID GeneralControlslimit of PV length in UserCalc record is 40 character for a PV
[6:38 PM] Dufresne, Eric
Hi Joe, I need help with a userCalc record. I had a userCalc in the past which worked. Recently the PV changed name, it is longer,
and I cannot get userCalc to add it. Is there a maximum length for a PV in the userCalc calculation? Will show what I mean in a picture.

&#8203;[6:40 PM] Dufresne, Eric
The prefix has just changed (from 8LAMBDA2) to DP_LAMBDA250K_1 and I want to write in the B field
userCalc6.png

[6:46 PM] Dufresne, Eric
In the above image with the new prefix, I cannot enter the full PV name on the B parameter, it clips to _R instead of _RBV. It is as if
someone put a maximum record length for the string at 39 characters.

[6:55 PM] Sullivan, Joseph
I'm afraid there is a maximum string length of 40 characters in that epics version.

The PV is 41 character long: DP_LAMBDA250K_1:cam1:NumImagesCounter_RBV Smile
It is the number of images taken by Area Detector.

We wanted to use this field to control some sample conditions.
We're using python instead of a userCalc as a result of this limitation.

ED

P.S. The MSD user created a python script to run this instead.
See Sep 2021 in spec_data.

Re: Sample stage


Hi all,

unfortunately, the way I have programmed this no longer works.
It is caused by a limit of the string length provided to a userCalc record.
This length is limited to 40 characters and my PV is 41 characters!
DP_LAMBDA250K_1:cam1:NumImagesCounter_RBV
It use to work because it started with 8LLAMBDA2.
The above PV is too long, but tells you the number of image collected by AD.
I am not reprogramming the detector and beamline at this time by changing the name of the detector.

What I am proposing instead would be to ignore this altogether and write
a simple python script that detects the time elapsed since it started, and from the time,
you can program a periodic function to drive the amplitude of the Keithley.

The voltage to apply is this PV: epics_put('8ide:2400:setVoltAO',30) (epics_put is a spec command).
An idea is to get the time of day from the IOC, a useful PV would be: 8idd:HEARTBEAT (time in seconds since start up)
8idd:TOD is a string with the calendar time of day called Current Time in the IOC status screen attached.
This can be grabbed for documentation. The heartbeat is the useful data to make an arbitrary
function generator.

You can make a python script with the frequency and amplitude and call it in spec
before the padtimescan call, making sure it is put in the background and stops
after Nseconds as programmed in the padtimescan call.

Hope this helps,
Eric
Attachment 1: UserCalc6-Bug.png
UserCalc6-Bug.png
Attachment 2: image.png
image.png
  805   Fri Sep 10 13:56:21 2021 Eric Dufresne8-ID GeneralComputers and Networkchanging from dynamic workspaces to static ones
Attachment 1: tweaks.png
tweaks.png
  804   Mon Aug 30 22:19:32 2021 Eric Dufresne8-ID-IDetectorsRigaku/EPICS bug with threashold adjustment
From: Narayanan, Suresh <[email protected]>
Sent: Wednesday, August 25, 2021 10:04 AM
To: Zhang, Qingteng <[email protected]>; Yasukazu Nakaye <[email protected]>; Hiroki Yoshida <[email protected]>
Cc: Takuto Sakumura <[email protected]>; Yasutaka Sakuma <[email protected]>;
Kazuyuki Matsushita <[email protected]>; Dufresne, Eric <[email protected]>; Lang, Keenan C. <[email protected]>
Subject: Re: Rigaku/epics

Dear Nakaye-san, Hiroki-san,

Out of curiosity, I tried changing the low energy threshold and I did not see
the number of non-zero pixels increase.
I lowered threshold to as low as 0.5 and saw not a single pixel increase.
Question is, is the field exposed to the library?

Keenan,

Can you check what happens in epics when the field is updated?

Thanks.

Suresh

RE: Rigaku/epics
Yasukazu Nakaye <[email protected]>
Mon 8/30/2021 7:00 PM
To: Narayanan, Suresh <[email protected]>; Zhang, Qingteng <[email protected]>; Lang, Keenan C. <[email protected]>; Hiroki Yoshida <[email protected]>
Cc: Takuto Sakumura <[email protected]>; Yasutaka Sakuma <[email protected]>; Kazuyuki Matsushita <[email protected]>; Dufresne, Eric <[email protected]>
Dear Suresh,

Sorry about my late response, but Hashimoto-san and Yoshida-san already informed QZ with instruction how to recover from the current situation.
Hashimoto-san found the cause of the problem and now we are working to fix the problem to update the server.
With current version, it looks like, if you set the threshold below 4 keV, server cannot find the files for
calculation and fails subsequently.

Yasukazu




Error logs

Re: Rigaku/epics
Narayanan, Suresh <[email protected]>
Fri 8/27/2021 12:40 PM
To: Zhang, Qingteng <[email protected]>; Yasukazu Nakaye <[email protected]>; Lang, Keenan C. <[email protected]>; Hiroki Yoshida <[email protected]>
Cc: Takuto Sakumura <[email protected]>; Yasutaka Sakuma <[email protected]>; Kazuyuki Matsushita <[email protected]>; Dufresne, Eric <[email protected]>
Dear Nakaye-san,

It seems like the entire controls is very fragile. I was using yesterday in the slow mode and it worked well.
Now the IOC wont take any images and see the “error: 33619976”

If these things keep happening, then we cannot run user operations. There are experiments that will fully rely
on this and we cannot get away with running Lambda for example. If slow mode does not work, we can get away
with Lambda. But if the whole thing fails, then we have to cancel the user experiment.

Why does this error keep happening? Is there a troubleshooting process that can be documented that we can
go through in steps? Usually, power cycle fixes most things but it is not clear here what to do.

Keenan, do you know what this error is?

Suresh

Re: Rigaku/epics
Narayanan, Suresh <[email protected]>
Mon 8/30/2021 9:11 PM
To: Yasukazu Nakaye <[email protected]>; Zhang, Qingteng <[email protected]>; Lang, Keenan C. <[email protected]>; Hiroki Yoshida <[email protected]>
Cc: Takuto Sakumura <[email protected]>; Yasutaka Sakuma <[email protected]>;
Kazuyuki Matsushita <[email protected]>; Dufresne, Eric <[email protected]>

Dear Nakaye,

Regarding log files, it would be nice if the most recent file always has the same name (maybe use a
sym link to always point to the current file), will be a good practice to manage things better.

Also, a database of error code and what it means could save a lot of trouble.

Suresh
  803   Thu Aug 26 12:17:29 2021 Eric Dufresne8-ID-ERepairsRecent X-ray BPM results from 8ID-E, dip in current on center
From: "Dufresne, Eric" <[email protected]>
Subject: Re: RFQ Request Argonne National Labs Quote # APS Dufresne m402 w. Coating & Adv
Date: August 26, 2021 at 11:39:35 AM EDT
To: Marychris Arcuri <[email protected]>

Hi Marychris,

can you please send this to your technical folks and ask them to get back to me?
The system I am using should still be under warranty.
I am having weird behavior with the inert gas unit, so I include
a report I wrote about the behavior.

Regards,
Eric

Fwd: RFQ Request Argonne National Labs Quote # APS Dufresne m402 w. Coating & Adv
Marychris Arcuri <[email protected]>
Thu 8/26/2021 10:48 AM
To: Martin, Benjamin W. <[email protected]>
Cc: Mark Katafiaz <[email protected]>; Dufresne, Eric <[email protected]>

1 attachments (2 MB)
July21-Report.pdf;

Hi Eric,
Your email has been received and I have included your technical point of contact on this email, Dr. Ben Martin.
Our team will review your report and reach out shortly.
Thank you,

Marychris Arcuri
Sydor Technologies
585.750.8578 | www.SydorTechnologies.com

RE: RFQ Request Argonne National Labs Quote # APS Dufresne m402 w. Coating & Adv
Valerie Fleischauer <[email protected]>
Thu 9/2/2021 12:52 PM
To: Dufresne, Eric <[email protected]>
Cc: Martin, Benjamin W. <[email protected]>; Marychris Arcuri <[email protected]>;
Matthew Zelazny <[email protected]>; Mark Katafiaz <[email protected]>

Hi Eric,

Thank you for sending along the report on the issue you are seeing with your diamond monitor. After looking through it we
have a few follow up questions (below) before making any conclusions. If it is easier and you are available for a call
next week to run through things please let us know.

  • The report mentions restarting the electrometer. What are the symptoms that drive the restart? i.e.
    software related crashes, or testing to see if this changes the diamond behavior
  • Has anything changed in the test setup or analysis since March? Or since the diamond was first installed?
  • Are all components in the inert gas purge UHP rated? Including tubing, regulator, etc.

Thank you,

Val

Val Fleischauer, Ph.D.
Technology Development Scientist, Sydor Technologies
1.585.278.1168 | www.SydorTechnologies.com

Re: RFQ Request Argonne National Labs Quote # APS Dufresne m402 w. Coating & Adv
Dufresne, Eric <[email protected]>
Fri 9/3/2021 3:43 PM
To: Fleischauer, Valerie E. <[email protected]>
Cc: Martin, Benjamin W. <[email protected]>; Marychris Arcuri <[email protected]>;
Matthew Zelazny <[email protected]>; Mark Katafiaz <[email protected]>

Valerie, all in cc,

I've found the electrometer not as stable as I would have expected. I've had to power cycle it sometimes few times a week to get
reliable signals. I don't have any error logs to report, but I'll make a table below of my observations in July.
Not very detailed, sorry.

Nothing changed since March, but some mechanical mounting. In March I changed motorized stages under it.

The BPM is always under N2 flow (High Purity Grade), and with the recommended flow.

Eric

7/13 Had to restart twice to remove large background currents of 11 uA. X-ray total current ~ 100nA.
7/19 Rebooted EM after end of experiment after noting dark currents were high.
7/23 BPM/EPICS not running. Did reset


Repair report link below:


https://anl.box.com/s/daq73cgioh6iqcxmcidoko4m9gvhvw5x

IEMR Report for M402 DBPM S/N: 017
Matthew Zelazny <[email protected]>
Wed 10/27/2021 9:41 AM
To: Dufresne, Eric <[email protected]>
Cc: Martin, Benjamin W. <[email protected]>; Marychris Arcuri <[email protected]>;
Fleischauer, Valerie E. <[email protected]>; Ziegler, Raymond <[email protected]>;
Mark Katafiaz <[email protected]>

Eric,

Attached is the IEMR report for you device. As you will see from the report the metallization of the sensor has been compromised.
Contamination in the operating environment is the suspected cause of the performance degradation.
Our recommendation is to replace the diamond and requalify the device.

Let us know how you wish to proceed.

Regards,

Matt

Confirmation of our conversation on DBPM
Rob Littlewood
Wed 11/24/2021 6:16 AM
To: Dufresne, Eric <[email protected]>
Cc: Mark Katafiaz ;Marychris Arcuri

Hi Eric,

This mail is just to confirm what we discussed last week; i.e. we will go ahead and replace the diamond in
your device for you for the lower fee of the quoted inspection and minor repair.

We will also provide updated information advising the use of He gas and let you know about any other pertinent
points to be observed when spec’ing other parts of the gas delivery system.

Kind Regards

Dr Rob Littlewood

Sydor Technologies
+44 (0)7917 567785
Skype: live:r.littlewood_1

Re: Confirmation of our conversation on DBPM
Dufresne, Eric <[email protected]>
Wed 11/24/2021 10:32 AM
To: Rob Littlewood <[email protected]>
Cc: Mark Katafiaz <[email protected]>; Marychris Arcuri <[email protected]>; Narayanan, Suresh <[email protected]>
Bcc: Ziegler, Raymond <[email protected]>

Thank you Rob,

you can count on me to get back to Sydor after the summer run of 2022 with any performance issue.
Helium makes sense too. I would appreciate getting an estimate of when I will get the repaired BPM back.

Happy Thanksgiving,

Eric



4/8/2022



Last week, I received the new Sydor X-ray BPM.
It is S/N 17, and its performance is attached in att. 2.
It was shipped around March 30, and arrived on site on Friday April 1.
I got it on my return to work from vacations.
Attachment 1: July21-Report.pdf
Attachment 2: SydorQAtesting-041222-4285_001.pdf
  802   Tue Aug 24 17:53:10 2021 Eric Dufresne8-ID-ISetupsnew scatterless pinhole from Incoatec for 8-ID-I

Suresh bought a new scatterless pinhole to act as a guard for 8-ID-I.
https://apps.inside.anl.gov/paris/req.jsp?reqNbr=G1-203053
See text below for details. Att. 1 is the pinhole holder.


[9:47 AM] Narayanan, Suresh
Jiang, Zhang Chu, Miaoqi is there any interest in scatterless-pinholes (not sure how scatter less they are, these are
from incoatec that are better than the ones we use, laser drilled and edges cleaned). I had talked to those people a while
back and he reached out now to close the case. While their apertures are bigger (100+ um), he said they offer 20-1000 um
Ta pinholes, 20 could end up a bit larger is my guess.

https://www.incoatec.de/scatex


Von: "Narayanan, Suresh" <[email protected]>
An: "[email protected]" <[email protected]>, "[email protected]"

<[email protected]>
Kopie: "Dufresne, Eric" <[email protected]>, "Jiang, Zhang"
<[email protected]>
Datum: 21.07.2021 16:54
Betreff: Re: Antwort: Re: Antwort: Re: Antwort: Re: Antwort: Re:
Antwort: Re: Antwort: Re: Antwort: Re: Antwort: Re: Antwort: RE: V0257
Antwort: Re: Antwort: Inquiry on scatterless pinholes

Hi,

I have been communicating with Andreas, seems like he is away. Could you
please modify the attached quote and replace the one there with the 55 um
that is in stock, we have decided to get that as we want to try sooner.
Thanks.

Suresh

Antwort: Re: Inquiry on scatterless pinholes
[email protected] <[email protected]>
Thu 7/22/2021 3:51 AM
To: Narayanan, Suresh <[email protected]>
Cc: Dufresne, Eric <[email protected]>; [email protected] <[email protected]>;
[email protected] <[email protected]>; Jiang, Zhang <[email protected]>

Hi Suresh,

we still have one on stock. Offical optically measured diameter 51 µm.
Depending on the energy of the x-rays it coulld appear a little bit larger
in your experiment. Is this okay?

I will send you the official offer later today.

Best regards
Jörg

Antwort: Re: Antwort: Re: Antwort: Re: Antwort: Drawing for Bruker holder V0257
[email protected] <[email protected]>
Tue 9/21/2021 7:13 AM
To: Narayanan, Suresh <[email protected]>
Cc: Dufresne, Eric <[email protected]>; [email protected] <[email protected]>
Hi Suresh,

The side that has Incoatec laser marked faces to the sample.

I hope I could help you.

Best regards,

Andreas

Antwort: Re: Antwort: Re: Antwort: Re: Antwort: Drawing for Bruker holder V0257
[email protected] <[email protected]>
Tue 9/21/2021 7:20 AM
To: Narayanan, Suresh <[email protected]>
Cc: Dufresne, Eric <[email protected]>; [email protected] <[email protected]>
Hi Suresh,

No problem. I try to help you.

You can see the pinhole in the holder. The side that you can see in the
holder is flat/polish. The cone side of the pinhole you can not see
because this side faces to the holder, this side is bonded to the holder.

The cone side of the holder faces to the sample.

Best regards,

Andreas

Performance is discussed in this research elog:
https://logbook.xray.aps.anl.gov/sector7/research/2225

Attachment 1: BrukerHolderA10C30.png
BrukerHolderA10C30.png
Attachment 2: pinhole.png
pinhole.png
  801   Thu Aug 19 12:36:09 2021 Eric Dufresne8-ID-EXPCS-APS-UNew Laser tracker for accurate detector distances


8/19/21
Called Chad Waldo Laser Trackers 816-207-8268
Zach Rode
Commercial Operations Manager - Midwest

Received in a wooden box set:
1) Spatial Analyzer USB key in New River Kinematics USB key
2) a USB dongle on a purple USB stick labelled Sentinel.
The latter is used with the software for license.

The software was installed with issue noted in att. 2.
You launch the software from the New Rivers Kinematics folder in the start menu
and it looks like this. We run it as 8ideuser.

From: Dufresne, Eric <[email protected]>
Sent: Thursday, September 2, 2021 9:12 AM
To: Fries, Mike <[email protected]>
Cc: Narayanan, Suresh <[email protected]>; Ziegler, Raymond <[email protected]>; Jansma, William G. <[email protected]>;
Highland, Matthew J. <[email protected]>; Wakefield, Kevin J <[email protected]>
Subject: New AT-403 Laser tracker from Leica tests on 8ID

Hi Mike,

for the APS-U 8-ID beamline, we've procured a new laser tracker from Leica, model AT-403.
Bill Jansma will help me with setting it up. It uses multiple class 1 lasers of various
wavelengths, and is classified as a class II laser.

Do I need any permission to start using it today?

Eric

RE: New AT-403 Laser tracker from Leica tests on 8ID
Fries, Mike <[email protected]>
Thu 9/2/2021 9:48 AM
To: Dufresne, Eric <[email protected]>
Cc: Narayanan, Suresh <[email protected]>; Ziegler, Raymond <[email protected]>; Jansma, William G. <[email protected]>;
Highland, Matthew J. <[email protected]>; Wakefield, Kevin J <[email protected]>

Hi Eric,

No issue with it being a class II, but we ask that it not be mounted at eye level. Thank you.

Mike Fries




9/2/2021

Summary, Bill set up the Laser Tracker today, reflecting on a 1.5" retroreflector
on the table. We installed additional software "Tracker Pilot" (see att. 5),
and in the afternoon, I connected to the APS wired network and
made a new connection with Tracker Pilot. I made a copy of the software on
my 8-ID-E folder. Mary registered it in IHW, and after rebooting it got the IP below.

Re: APS Dead Network Node(s)
Westbrook, Mary L. <[email protected]>
Thu 9/2/2021 5:14 PM
To: Dufresne, Eric <[email protected]>

K, Eric,
I have added this device to IHW and assigned the IP address below:
at403-8ide.xray.aps.anl.gov
164.54.116.53

thanks, mary


9/7/2021 Battery charging


Power Supply Questions for AT403 Laser Tracker
FORREST Phillip <[email protected]>
Tue 9/7/2021 7:35 AM
To: Dufresne, Eric <[email protected]>

1 attachments (422 KB)
Power Supply Options.pdf;

Eric,
I checked with my colleague Shawn this morning on the AT403 Battery Questions you asked on Friday afternoon, here is Shawn’s response.

for the 960 and the 40x trackers the batteries are not charged when plugged in with power cable as well-the batteries are only for back up
power or when power is unplugged for a move or to be used when in an area that power is not available. The batteries are only charged when
on the charging station for the AT 40x trackers. For the AT 960 systems the batteries are charged when power cable is plugged into the
battery but will not charge if they are only connected to the controller-does that answer your questions

Let me know if you need anything else,

Phillip Forrest
Technical Support
Hexagon Manufacturing Intelligence
Support.HexagonMI.com
[email protected]

1-800-343-7933
Hexagon Detroit Solutions Center
46444 Hexagon Way
Novi, MI 46480
USA


3/7/2022 Spatial Analyzer license



Spatial Analyzer
Jansma, William G. <[email protected]>
Sun 3/6/2022 11:25 AM
To: Highland, Matthew J. <[email protected]>; Dufresne, Eric <[email protected]>

Hi Matt; Eric,

Would you please locate the 5 digit license dongle number for your seat of Spatial Analyzer software and send it to me?
It is located on the back of the USB license key...
I received notice for renewal and there is a number I am not familiar with - thinking it might possibly be your license.
The numbers were 10518 and 10519 - maybe you have one of them?
Kindly let me know...

Thank You,

Bill

William G. Jansma
Survey & Alignment Engineer
Survey & Alignment Technical Lead
Photon Sciences
[email protected]
(630) 252-9353 Office
(708) 218-2264 Mobile

Re: Spatial Analyzer
Dufresne, Eric <[email protected]>
Mon 3/7/2022 11:59 AM
To: Jansma, William G. <[email protected]>
Cc: Highland, Matthew J. <[email protected]>

Bill,

I confirm the number 10518 is for 8ID-E.
Thanks for highlighting this, I am making a note for myself.

Eric

Re: Spatial Analyzer
Jansma, William G. <[email protected]>
Mon 3/7/2022 12:47 PM
To: Dufresne, Eric <[email protected]>
Cc: Highland, Matthew J. <[email protected]>

Hi Eric,

I think there has been a mix-up in who got which license.
License #10518 is the Pro version of Spatial Analyzer, and #10519 is the Ultimate version.
We are in possession if #10519, but I recall that we ordered the Pro version.
Can you confirm whether your order included the Ultimate version? This should be listed in your original invoice from Hexagon.
If so, we will need to exchange the license keys... Sorry for the confusion - this just came to my attention.

Kind regards,

Bill

Re: Spatial Analyzer
Dufresne, Eric <[email protected]>
Mon 3/7/2022 1:07 PM
To: Jansma, William G. <[email protected]>
Cc: Highland, Matthew J. <[email protected]>

Hi Bill,

we ordered the Ultimate as I checked in our invoice.
I'm onsite regularly, available until 3 today.

Cheers,
Eric

P.S. Matt- invoice is https://logbook.xray.aps.anl.gov/8id/8-ID/801/1 here (VPN needed)
Attachment 1: LaserTrackerPackingSlip3719_001.pdf
Attachment 2: NewRiverInstallationIssuesNoted.png
NewRiverInstallationIssuesNoted.png
Attachment 3: SpatialAnalyzer.png
SpatialAnalyzer.png
Attachment 4: 3754_001.pdf
Attachment 5: TrackerPilot3p4p1p318.png
TrackerPilot3p4p1p318.png
Attachment 6: PowerSupplyOptions.pdf
  800   Sun Aug 1 19:27:47 2021 Eric Dufresne8-ID-EComputers and Networkcomplete freezing of NX for 8ID-E.
See also controls issue with s8idigui and matlab
https://logbook.xray.aps.anl.gov/8id/controls/126 (several IOC crashes because of operator error)

[Chat with SN 10:25 AM] Dufresne, Eric

Some progress has been made on freezing NX for 8ID-E. It looks as though specr causes the full workstation to lock down
when you try to overlay data from multiple scans. Note the image below. I opened up pymca for them and showed Lingjia
how to plot multiple scans with it. Killing matlab doesn't help, did a full reboot.
I tried restarting NX but this failed for me or the beamline account.

Also yesterday decided to use virtual as a test. Still crashed at 9am this morning. Virtual had similar speed as physical in Sweeden,
but lower resolution. It is nicer so I'll stay with the physical display.

[matlabCrash-9am-080121.png] (https://argonnedoe-my.sharepoint.com/personal/dufresne_anl_gov/Documents/Microsoft\ Teams\ Chat\ Files/matlabCrash-9am-080121.png)
Attachment 1: matlabCrash-9am-080121.png
matlabCrash-9am-080121.png
  799   Tue Jul 20 10:46:58 2021 Eric Dufresne8-ID GeneralBeamline HardwareVirtual Tour Eric and Qingteng gave of 8-ID-A, D, E and I, plus lab tour all filmed by Qingteng
[4/29 10:44 PM] Zhang, Qingteng
Dear All, here's the Box link to the Virtual Tour Eric gave to 8-ID-A, D, E and I, plus my
short tour of the chem lab featuring the use of QNW and the loading of the cell, totaling ~43 min.
https://anl.box.com/s/sp99nqvc72wyi6b7kan8h7rlvm444wcn

[4/29 10:46 PM] Zhang, Qingteng
We had to shoot it twice because of my technical mistake the first time (Never rotate camera in the middle
of the recording if you want to view it on computers later). I am very grateful for Dufresne, Eric's
understanding and his knowledge and expertise throughout the entire process. It was a great learning experience for me

[4/29 10:48 PM] Zhang, Qingteng
There's an extra ~ 13 min video Eric recorded in which he went through a bit more details on the alignment
at the A hutch. It was very informative for me and may be of help to Peco, Ash and Miaoqi as well. I highly
recommend this series. Narayanan, Suresh Maybe we could put this in #general on both 8-ID-I and 8-ID-E Slack?
  798   Thu Jul 15 18:02:42 2021 Eric Dufresne8-ID-EControlsBasic epics for Keithley 2400 controls, 6ID Keithley 2410 from 6-ID

Phil Ryan and I asked Keanan Lang to set up a second Keithley 2400s series from 6ID in parallel
with ours because we were using it and I care about keeping it as is.
This was implemented 7/16/21 at 6pm.
Their Keithley 2410 operates up to 400V, ours only to 211 V.
It is set at 9600 bauds instead of ours at 38400 Bauds.
caQtDM screens are shown below.
The *idn? command also works with Keithley!
This commands is recognized by Lakeshore.


Re: Keithley 2410 at 8-ID
Choi, Yongseong <[email protected]>
Fri 7/16/2021 6:48 PM
To: Lang, Keenan C. <[email protected]>; Dufresne, Eric <[email protected]>;
Ryan, Philip J. <[email protected]>; Walsh, Thomas R. <[email protected]>
Hi Keenan,

Thanks for setting up the database.

We rebooted the crate. Now all the PV values for the Keithely 2410 are accessible (from caget).
However, the ui screen for the 2410 unit has blank lines (below is the new screen). It looks like there is a
typo somewhere in caQDTM so that the values are not being displayed properly. Could you check the ui file for 2410?

Thanks.
Yong

Re: Keithley 2410 at 8-ID
Lang, Keenan C. <[email protected]>
Fri 7/16/2021 8:55 PM
To: Choi, Yongseong <[email protected]>; Dufresne, Eric <[email protected]>;
Ryan, Philip J. <[email protected]>; Walsh, Thomas R. <[email protected]>

Yep, looks like I missed a comma. You'll have to reload the 8ide.ui screen
but then launching the 2410 screen should work.


epics_put('8ide:2400:setVoltAO',30)
padtimescan 300 2


pepper% caput 8ide:2410:enableBO 1
Old : 8ide:2410:enableBO Off
New : 8ide:2410:enableBO On
pepper% caget 8ide:2410:measResAI
8ide:2410:measResAI 0



Quote:
I made a function generator out of user Calc records to generate a square wave.
The time axis is connected to the AreaDetector Image Counter.
Because the expression is too long, it connects userCalc7 to userCalc6.
Note that the output of userCalc6 is read in variable A in userCalc7.
The calculation can only be as long as I have in userCalc6 so I had to nest two userCalcs.
In this attachment showing a test, userCalc6 goes from 0 to 1 with a period of NumImages/10 = 60 files, the switch occurs
every 30 files and it goes from 0, 0.5 and 1 in three files.
I would try to keep in the number of periods even like 2,4...
It drives the Keithley Source meter https://logbook.xray.aps.anl.gov/8id/8-ID/528
so this is convenient for higher excitation voltages.

An earlier version is shown in att. 3 for userCalc5.
It did only on 50% off 50%.

My users also made a sinusoidal version in att. 4.

ED
Attachment 1: asyn-8ide.png
asyn-8ide.png
Attachment 2: asyn-18-8ide-serial.png
asyn-18-8ide-serial.png
Attachment 3: asyn-18-8ide.png
asyn-18-8ide.png
Attachment 4: Keithley2410-6ID.png
Keithley2410-6ID.png
  797   Tue Jul 13 18:31:22 2021 Eric Dufresne8-ID-EDetectorsLambda250k upgrade to 24 kHz 1 bit, integration like 8ID-I with softglue
[6/25 7:38 PM] Narayanan, Suresh
lambda250k with all the modes with and without trigger has been tested quite thoroughly from spec.
All the shutter functionality for detector control is like 8idi now, can run with or without shutter
for fast and slow measurements, maybe helpful for some rad sensitive samples, documented it
reasonably well in paper, not as good as your docs

Att. 1 is from Suresh detailing the Softglue work done in May.

[7/113/21 10:04 PM] Narayanan, Suresh
Sure. Cabling was a lousy job that day with the DP person, trying to do it when i was out the door.
IOC is running on the DP PC on the rack.
[7/13/21 10:05 PM] Narayanan, Suresh
One thing I did not check was the physical shutter opening. The status of the fast shutter.ui
screen works like before so I assume it should be good.



New softglue: https://logbook.xray.aps.anl.gov/8id/8-ID/787
PO/Upgrade notes from Suresh, Xspectrum and DP staff: https://logbook.xray.aps.anl.gov/8id/8-ID/773


See new spec macros:

xpcs_ad_nofpga_lambda
use_shutter_mode_regular
use_shutter_mode_burst


From batches/2021/May2021 lambda.do



##Usage: padtimescan ndata xtime [period]
# For shuttered mode, choose LAMBDA_OPERATING_MODE = 1 and EXT_TRIGGER=2
# For ext. trigger mode 0, 1, choose LAMBDA_OPERATING_MODE = 0
# zero deadtime (ZDT)

LAMBDA_OPERATING_MODE = 3; ##0-1 bit ZDT, 1-6-bit ZDT, 2 - 12 bit ZDT, 3 -24-bit
mode
EXT_TRIGGER=2

att 0
ARun_number++
batch_name=sprintf("A%03d_delete",ARun_number)
#padtimescan 50000 1/24000.0
padtimescan 50 .1 .55
p INFOFILE

EPICS screen now on 8ID-E Engineering screen



Click on AreaDetector in att. 2.
The full screen appears as shown in att. 3.
Pushing the "less button" at the bottom of the screen gives a smaller, narrower window.


How to start IOC



xspadmin@xspserver:~/bin$
xspadmin@xspserver:~/bin$ ./DP_LAMBDA250K_1.sh status
DP_LAMBDA250K_1 is running (pid=13422) in a screen session (pid=13421)
xspadmin@xspserver:~/bin$ ./DP_LAMBDA250K_1.sh stop
Stopping DP_LAMBDA250K_1 (pid=13422)

xspadmin@xspserver:~/bin$ ./DP_LAMBDA250K_1.sh
Usage: DP_LAMBDA250K_1.sh {start|stop|restart|status|console|run|medm|caqtdm}

Additional options:
DP_LAMBDA250K_1.sh start {screen|procServ|ps|shell}

xspadmin@xspserver:~/bin$ ./DP_LAMBDA250K_1.sh status
DP_LAMBDA250K_1 is not running
Attachment 1: WAXPCS_softglue_docs.pdf
Attachment 2: NewCaQtDMscreen-Lambda250k.png
NewCaQtDMscreen-Lambda250k.png
Attachment 3: LambdaBase-May21.png
LambdaBase-May21.png
Attachment 4: 8idMZ1-SG.png
8idMZ1-SG.png
Attachment 5: 8idMZ3-SG.png
8idMZ3-SG.png
  796   Tue Jul 13 15:21:29 2021 Eric Dufresne8-ID-EControls2020-2021 Changes to EPICS controls in 8ID-E (X-BPM, )

1) To start Newport XPS
https://logbook.xray.aps.anl.gov/8id/controls/120 (Full installation)
IOC control:
ioc8idexps [start, stop, console]
fyi: The XPS 8IDE IOC is running out of /net/s8iddserv/xorApps/epics/synApps_6_1/ioc/8idexps/iocBoot/ioc8idexps

Screens:
https://logbook.xray.aps.anl.gov/8id/controls/120/5

See below for details from elog 120.
glass% nslookup 164.54.116.122
122.116.54.164.in-addr.arpa name = s8xps2.xray.aps.anl.gov.

glass% ping s8xps2
PING s8xps2.xray.aps.anl.gov (164.54.116.122) 56(84) bytes of data Works.


2) BPM Sydor

/net/s8iddserv/xorApps/epics/synApps_6_0/ioc/8ideBPM/iocBoot/iocbpm/8ideBPM.sh css

/net/s8iddserv/xorApps/epics/synApps_6_0/ioc/8ideBPM/iocBoot/iocbpm/8ideBPM.sh usage
Usage: 8ideBPM.sh {start|stop|restart|status|console|run|css}

[12/3/20 1:22 PM] Peterson, Kevin M.
start will start the IOC in a screen session, run will run the IOC on the command line outside of screen.

[12/3/20 1:22 PM] Peterson, Kevin M.
If you run 8ideBPM.sh with start, and the IOC doesn't appear to be running, use the run argument and you'll see the error.

To summarize, you turn on the unit, then start the ioc with 8ideBPM.sh run.
Then you follow with 8ideBPM.sh css to start CSS control environment.

Only one instance of this process is allowed.
  795   Wed Jun 30 10:31:52 2021 Eric Dufresne8-ID-AProblemSolution to autoshutter not opening shutter in 8-ID-A
[6/24 2:31 PM] From Suresh
Thomas, Joe, the 8-ID autoshutter control for the past month or more, even maybe last cycle,
I enable on Monday afternoon and find that before next morning it is disabled again.
Could you please take a look?


[8:34 AM] From Joe Sullivan
Suresh,
the 'Lease Date' has expired. This does a Disable when the permits change states.
Bump up the lease date the next time the A-Shutter is closed.

[10:38 AM] Walsh, Thomas R.
What is the reason for the lease Date?

&#8203;[2:15 PM] Narayanan, Suresh
to make people remember that such an auto tool exists that can keep opening the shutter even if you dont want to.
I am serious, incident happened at 33id 8 years ago that kept opening the shutter while beam was burning through a bellow

&#8203;[2:15 PM] Narayanan, Suresh
based on that incident, lease feature was added by Joe and also the auto disable on Monday was added as well

7/6/2021 Update
Changed lease date to 11/30/2022... Should fix the problem Suresh mentioned above.
Attachment 1: ScreenShot_2021-06-30at8h28m07_AM.png
ScreenShot_2021-06-30at8h28m07_AM.png
Attachment 2: autoShutter-NewLease.png
autoShutter-NewLease.png
  794   Wed Jun 23 19:48:15 2021 Eric Dufresne8-ID GeneralComputers and Networkwww.xray.aps.anl.gov and how to make your own web page

A few of us on 7ID have web pages located on our /home/beams/ directory.
Dave Wallis has made it easy recently to view these web pages with
a personnal web service:

https://www.xray.aps.anl.gov/~dufresne/

My files are in ~/public_html which means
/home/beams/DUFRESNE/public_html
and the top file displayed on my web page
/home/beams/DUFRESME/public_html/index.html

Could be also index.htm if you save it with windows.

I find it sometimes very helpful to share files quickly so I have
a pub directory:
/home/beams/DUFRESNE/public_html/8ID/8ID-E/pub

Others have made web pages in the past at 7ID:
https://7id.xray.aps.anl.gov/
https://7id.xray.aps.anl.gov/~dwalko/
https://7id.xray.aps.anl.gov/~wen/

This is FYI.

ED